Chapter 8
Options and Upgrades
1062788
BiPAP Vision Service Manual © Respironics, Inc.
8-11
8.3
PC/MC/DC
Upgrade
Installation
This section describes how to install these upgrades for ventilator s/n 105999
and below:
•
PC/MC/DC three-board upgrade, PAV (P/N 104707GL)
•
PC/MC/DC three-board upgrade, non-PAV (P/N 1004714GL)
CAUTION:
Perform this repair according to proper ESD guidelines.
NOTE:
Record the total operating hours to enter onto the ventilator when the
upgrade is complete.
Required equipment:
•
Ribbon cable, test (P/N 1004699)
•
Test cable (P/N 582161)
•
Test cable (P/N 1004823)
Fuse, 100-120 V, 3.5 A
(P/N 1000749)
Fuse, 100-120 V, 4.0 A (P/N 1041196)
If current fan muffler is labeled with 3.5-A rating, also
install 4.0-A rating circulation fan muffler
(P/N 1041193).
MCS board (P/N 582140)
MC upgrade kit, non-PAV (P/N 1004713)
MC upgrade kit, PAV (P/N 1000356)
OM assembly,
French (P/N 582254)
Spanish (P/N 582255)
Italian (P/N 1003547)
German (P/N 582220)
OM assembly, international (P/N 1004977)
PAV/T option kit (P/N 1000747)
EPROM kit with PAV for s/n 105999 and below not
previously upgraded (P/N 1003254)
Power line filter (P/N 1000745)
AC inlet (P/N 582138)
Pressure airflow subsystem (PAS)
board (P/N 582146)
PC upgrade kit, non-PAV (P/N 1004714GL)
PC upgrade kit, PAV (P/N 104707GL)
Touch pad
French (P/N 582257)
Spanish (P/N 582256)
Touch pad, international (P/N 1004712)
Table 8-2: Obsolete Repair Kits
Obsolete kit
Substitution repair kit