Periodic Maintenance
4.6.3
Verifying the Charge
At least two hours is required to charge a fully discharged battery to a voltage
to a level that does not activate a low voltage alarm. After 2 hours, the
ventilator can be operated and continues to trickle-charge the battery during
operation.
1. Press Monitoring to begin operation.
2. Wait 2 minutes to determine if the
Check Vent
alarm activates and an
error code 205 is displayed in the Error Message screen. If not, then
the unit is ready for use.