Chapter 5
5-8
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Component Removal/Installation
5.5 Fuses
Follow these steps to replace ventilator fuses (Figure 5-7, Figure 5-8):
1. Turn the ventilator off and disconnect the power cord from the back of
the device and the wall outlet.
2. Gently pry open the top of the fuse holder door. The door hinges
downward.
Figure 5-7: Opening the Fuse Door
Ventilator s/n
100500 and above
•
Use P/N 1000749 (100 and 120 VAC)
•
Use P/N 1000750 (230 and 240 VAC)
Ventilator s/n
100499 and below
•
Use P/N 582100 (115 VAC)
•
Use P/N 1000750 (220 and 240 VAC)
Gently pry fuse
door here