Configuring the measurement setup
R&S
®
AREG800A
77
User Manual 1179.3615.02 ─ 01
Angle Frontend to Sensor
Sets the angle of the field-of-view between frontend and sensor.
Frontend
Sensor
IF
Radar Echo
Signal Gener
d
1
2
1 = Distance frontend sensor
2 = "Angle Frontend to Sensor"
Remote command:
[:SOURce<hw>]:AREGenerator:FRONtend:TRX<ch>|QAT<ch>:ATS
Rotation Frontend to Sensor
Sets the angle of the field-of-view between frontend and sensor.
Remote command:
[:SOURce<hw>]:AREGenerator:FRONtend:TRX<ch>|QAT<ch>:RTS
Use Custom Antenna
Activates configuration of a custom antenna assembly.
If activated, you can configure the antenna gain that is specific for this antenna assem-
bly. Configure the antenna gain for the transmitting antenna (TX) and for the receiving
antenna (RX):
●
●
Remote command:
[:SOURce<hw>]:AREGenerator:FRONtend:TRX<ch>:ANTenna:CUSTom[:
STATe]
AREG Antenna Gain TX
Displays the antenna gain of transmitting antenna (TX) that is mounted at the
R&S
AREG800A.
Configuring a custom gain value, requires "Use Custom Antenna > On". Set the result-
ing total TX antenna gain according to your antenna assembly mounted at the
R&S
AREG800A.
Remote command:
[:SOURce<hw>]:AREGenerator:FRONtend:TRX<ch>:ANTenna:GAIN:TX
Frontend configuration