Remote control commands
R&S
®
AREG800A
259
User Manual 1179.3615.02 ─ 01
The dialog ID is also required to define user key actions.
See
Chapter 9.2.4, "How to assign actions to the [User] key"
// Query the dialog IDs of all open dialogs
:DISPlay:DIALog:ID?
// CEUtraDLGenSetDlg,_,$A DlgKeyRf_Rosc
// Open and close dialogs via remote control
:DISPlay:DIALog:OPEN "CEUtraDLGenSetDlg,_,$A"
:DISPlay:DIALog:OPEN "DlgKeyRf_Rosc"
:DISPlay:DIALog:CLOSe "DlgKeyRf_Rosc"
:DISPlay:DIALog:CLOSe:ALL
............................................................................................... 259
.................................................................................................259
.........................................................................................260
............................................................................................... 260
..................................................................................... 260
....................................................................................261
..............................................................................................261
................................................................................................ 262
..........................................................................................263
:DISPlay:PSAVe:HOLDoff
<HoldoffTimeMin>
Sets the wait time for the screen saver mode of the display.
Parameters:
<HoldoffTimeMin>
integer
Range:
1 to 60
*RST:
n.a. (factory preset: 10)
Default unit: minute
Example:
Example "Activating screen saver mode and display update"
Manual operation:
See
:DISPlay:PSAVe[:STATe]
<State>
Activates the screen saver mode of the display.
We recommend that you use this mode to protect the display, if you operate the instru-
ment in remote control.
To define the wait time, use the command
.
DISPlay subsystem