General instrument functions
R&S
®
AREG800A
159
User Manual 1179.3615.02 ─ 01
8. On the source instrument, select "System Config > Setup > Instrument Assembly >
Manage License Keys > Manage License Keys > Export Deactivation Response to
File".
9. In the browser, go to "Manage Licenses > Move Portable License > Install Deacti-
vation Key (from file)".
10. Enter the deactivation response of the instrument.
The license is deactivated for the source instrument.
11. In the "Manage Licenses > Move Portable License", go to step "Create License" to
generate a license key for this portable option and the selected target instrument.
a) Download the license key as a file.
b) Transfer it to the target instrument.
12. In the target instrument, select "System Config > Setup > Instrument Assembly >
Manage License Keys > New License > Import License Keys from File".
13. Select the created license key file.
The portable option is installed on the target instrument.
9.4
Using the security settings
The protection function of the R&S
AREG800A offers several levels to activate particu-
lar functions like self-test or tests for service purposes specifically.
Protection
The five protection levels are automatically active on startup, the protection levels, that
means all protected functions are locked.
To unlock a protection level:
► In the "System Config > Setup > Security > Protection", enter the correct password.
To lock a protection level:
► Clear the corresponding checkbox.
Protection levels
The following functions are protected in the respective levels:
●
Protection level 1
Protects against accidental changes, like, for example, the clock and date, several
internal adjustments functions and the self-test, as well as network settings or the
instrument hostname.
You can access this level with the password 123456.
●
Protection level 2
Using the security settings