Remote control commands
R&S
®
AREG800A
296
User Manual 1179.3615.02 ─ 01
AUTO
Automatically determines the strings.
USER
User-defined strings can be selected.
*RST:
n.a. (factory preset: AUTO)
Example:
SYST:IDEN AUTO
Automatically assigns the OPT and IDN strings according to the
selected instrument language.
Manual operation:
See
:SYSTem:IDENtification:PRESet
Sets the
*IDN
and
*OPT
strings in user defined mode to default values.
Example:
SYST:IDEN USER
SYST:IDEN:PRES
Usage:
Event
Manual operation:
See
:SYSTem:IRESponse
<IdnResponse>
Defines the user defined identification string for
*IDN
.
Note
: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command
SYST:IRES
is discarded.
Parameters:
<IdnResponse>
string
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:IRES "Test Device"
// Defines identification string 'test device'
*IDN?
// Response: 'test device'
Manual operation:
See
:SYSTem:ORESponse
<OResponse>
Defines the user defined response string for
*OPT
.
Note
: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command
SYST:ORES
is discarded.
Parameters:
<OResponse>
string
SYSTem subsystem