Maintenance, storage and disposal
R&S
®
AREG800A
378
User Manual 1179.3615.02 ─ 01
Temperature Offset
Displays the temperature difference, comparing the temperature of the last adjustment
to the current instrument temperature.
A green checkmark indícates that the offset is within the permitted range. If the temper-
ature deviates more than
±
5
K, the instrument indicates a warning icon.
Remote command:
Information
Displays information to the current adjustment state.
Remote command:
:CALibration<hw>:ALL:INFormation?
Continue Adjustment on Error
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
:CALibration<hw>:CONTinueonerror
14.3.1.2
How to use the internal adjustments
Deciding whether to run internal adjustments
1. Select "System Config > Setup > Internal Adjustment".
2. Observe the status and color indication in the section "Since Last Full Adjustment".
Green: Internal adjustments are not required.
Red: Internal adjustments are required. Observe also the indication in the "Informa-
tion" field.
3. We recommend that you run internal adjustments in the following cases:
● Before starting any application, that requires a maximum of level accuracy
● When a long period of time has passed since the last adjustments
● If the ambient temperature of the instrument significantly differs from the one of
the last adjustments
4. Proceed as described in
"Running internal adjustments"
Running internal adjustments
1.
NOTICE!
Adjustments can be invalid if performed when the instrument is not
warmed-up.
Wait until the instrument has reached its operating temperature before you start the
adjustment procedure.
The warm-up time is up to 30 minutes.
Performing maintenance tasks