Getting started
R&S
®
AREG800A
19
User Manual 1179.3615.02 ─ 01
Prevent overheating as follows:
● Keep a minimum distance of 10
cm between the fan openings of the product
and any object in the vicinity.
● Do not place the product next to heat-generating equipment such as radiators
or other products.
3.1.4.2
Mounting the R&S
AREG800A in a rack
To prepare the rack
1. Observe the requirements and instructions in
2.
NOTICE!
Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S
AREG800A in a rack
1. Use an adapter kit that fits the dimensions of the R&S
AREG800A to prepare the
instrument for rack mounting.
a) Order the rack adapter kit designed for the R&S
AREG800A. For the order
number, see data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2.
WARNING!
The R&S
AREG800A can be heavy, e.g., if fully equipped. Use a lifting
"Lifting and carrying the product"
Lift the R&S
AREG800A to shelf height.
3. Grab the handles at the front and push the R&S
AREG800A onto the shelf until the
rack brackets fit closely to the rack.
4. Tighten all screws at the rack brackets with a tightening torque of 1.2
Nm to secure
the R&S
AREG800A in the rack.
To unmount the R&S
AREG800A from a rack
1. Loosen the screws at the rack brackets.
2.
WARNING!
The R&S
AREG800A can be heavy, e.g., if fully equipped. Use a lifting
"Lifting and carrying the product"
Bring the lifting equipment to shelf height.
3. Remove the R&S
AREG800A from the rack.
4. If placing the R&S
AREG800A on a bench top again, unmount the adapter kit from
the R&S
AREG800A. Follow the instructions provided with the adapter kit.