The measured value can be calibrated by changing the cell constant to the actual conductivity
value (arrow keys).
Select “install. factor setting?” in the setting menu, note the set installation factor and then set
to 1.00
Select the fi rst menu option in the “calibration cc” setting menu and press Enter
Press Enter when the measured value under “LF” is stable
“please wait!” appears for a few seconds, then the DULCOMETER
®
D1C displays the current
calculated cell constant
The new cell constant is accepted when you press Enter to exit the last menu option (“calibra-
tion cc: - cc = ...”). Exit this menu option with the back key if you do not want to accept the
new cell constant.
Select “install. factor setting?” in the setting menu and restore the installation factor noted
earlier.
Next, restore the temperature coeffi cient
of the water sample used
It is highly recommended that you check the conductivity of the calibration solution with
suitable reference metering equipment to be sure that the conductivity of the calibration
solution has not changed due to carry-over of liquid or evaporation.
Inductive Conductivity
Conductivity Calibration
Measurement
Parameters
100
μ
S/cm
Permanent display 2
only with control
(w = setpoint)
D1C2-cond.-011-GB
Permanent display 1
mea val.
fd. fwd.
reg. val.
70 %
59 %
100
μ
S/cm
1000
μ
S/cm
w:
Positive values of setting variable: Increase conductivity
Negative values of setting variable: Reduce conductivity
D1C2-cond.-040-GB
calibration cc
probe in buffer!
accept value?
1538
μ
S/cm 8.54
calibration cc ?
PAR1
calibration cc
cc: 8.54 /cm
cond: 100
μ
S/cm
calibration cc
cc: 8.50 /cm
please wait!
Cell Constant Calibration (cc)
D1C2-cond.-030-GB
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
mr = meas. range
cc = cell constant
2000
μ
S/cm
cc :
mr :
1106
μ
S/cm 25,0
°
C
mr 2 : 2000
μ
S/cm
cc 2 : 8.54 /cm
1106
μ
S/cm 25,0
°
C
8.54/cm