Pr
o
Minent
®
Page 119
pH/Redox (ORP)
Operating Menu / Description
Permanent
displays
The permanent displays 1 to 4 serve information on fault messages/causes as well as on op er at-
ing values/settings.
8.10.2
DULCOTEST
®
Redox (ORP) Sensor RHER-Pt-SE and DULCOTEST
®
Sensor PHEP or PHER for pH, Commissioning
8.10.2.1
DULCOTEST
®
Redox (ORP) Sensor RHER-Pt-SE and DULCOTEST
®
Sensor PHEP or PHER for pH, Assembly
Assembly
Redox (ORP) Sensor
Before putting into operation, the protective cap or case must be removed from the glass or
metal electrode. The glass or metal electrode must be free from oil, grease and other contamina-
tions. Likewise, the diaphragm of the reference electrode must be free from scale deposits, dirt
or crystallized matter. For these reasons, electrodes should never be touched by hand.
Remove the blanking plug of the respective measuring module at the fi ttings of the measuring/
control station DULCOTROL
®
and install the DULCOTEST
®
RHER-Pt-SE sensor for redox (ORP).
pH
Sensor
Before putting into operation, the protective cap or case must be removed from the glass or
metal electrode. The glass or metal electrode must be free from oil, grease and other contamina-
tions. Likewise, the diaphragm of the reference electrode must be free from scale deposits, dirt
or crystallized matter. For these reasons, electrodes should never be touched by hand.
Remove the blanking plug of the respective measuring module at the fi ttings of the measuring/
control station DULCOTROL
®
and install the DULCOTEST
®
PHEP or PHER sensor for pH.
8.10.2.2
DULCOTEST
®
Redox (ORP) Sensor RHER-Pt-SE and DULCOTEST
®
Sensor PHEP or PHER for pH, Electrical Installation
Installation
Redox (ORP) Sensor
Connect the DULCOTEST
®
RHER-Pt-SE sensor for redox (ORP) with the marked cable of the
respective controller DULCOMETER
®
D2C.
pH
Sensor
Connect the DULCOTEST
®
PHEP or PHER sensor for pH with the marked cable of the respective
controller DULCOMETER
®
D2C.
7.20
750
pH
mV
value 1 7.20 pH
value 2 750 mV
stop-key
value 2 750 mV
set point 750 mV
ctrlout
0 %
value 1 7.00 pH
set point 7.20 pH
ctrlout
0 %
D2C2-pH-Re-MT-1-D
BA_DR_002_02_09_GB_Cooling.indd 119
BA_DR_002_02_09_GB_Cooling.indd 119
24.02.2009 11:56:11 Uhr
24.02.2009 11:56:11 Uhr