Pr
o
Minent
®
Page 53
Operation
Note text
Symbol
Effect
Alarm with ack- Remarks
Remedy
on metering on control
nowledgement
Pause contact
Pause
Stop
Stop
No/Yes*
No further fault check
–
Pause/Hold
PI-part
frozen
Stop button
Stop
Stop
Stop
No
Relay drops out
–
During calibration sensor
Basic load
Stop in
No
No error processing of
–
+ Feed
complete
measured variable
forward
operating
control
menu
Sensor slope too low
Slope low
25% > sensor of slope
Check sensor,
Sensor slope too high
Slope high
Basic load
Stop
No
> 300% norm slope
replace if necessary
H
2
O
2
-value <2 % measuring range
H
2
O
2
too low
< 2 % from meas. range
Zero point
Zero point low
Basic load
Stop
No
Signal < 3 mA
Check sensor/cable
Zero point high
Signal > 5 mA
Repeat calibration
in H
2
O
2
-free water
During servomotor setting
Position feed back wrong
Direction check
Without correct
Check connection of relay,
Upper position <40 % max. value
Final value small
adjustment the last
potentiometer
Lower position >30 % range
Final value big
valid values are still used
Adjust the operation region
of the servomotor correctly
*dependent on whether “Alarm on” or “Alarm off” is set in “General settings”
Sensor:
Troubleshooting
Troubleshooting includes the complete measuring station. This consists of
1)
Measuring/control
equipment
2) Electrical cable and connections
3) In-line probe housing and hydraulic connections
4)
Sensor
Possible causes for faults shown in the table below mainly refer to the sensor. Before beginning
to look for any faults, you should ensure that all operating instructions have been carried out in
accordance with the Technical data:
a)
H
2
O
2
content in accordance with the measurement range
b) Sample water temperature 0 - 50 °C and constant
c) Flow rate 30 - 60 l/h
You can use the sensor simulator (DULCOMETER
®
simulator order no. 1004042) to locate the
fault in the measuring and control system. The operating instructions manual for the
DULCOMETER
®
D1C measured value H
2
O
2
gives full details on how to locate a fault in the
measuring and control equipment.
Where there are major discrepancies between the measured values of the sensor and the meas-
ured value of the reference methods, you should fi rst consider all possible faults relating to the
reference methods. You should repeat the reference measurement several times if necessary.
H
2
O
2
Fault
Sensor cannot be
calibrated and measured
value of the sensor is
greater
than the reference
measurement
Sensor cannot be
calibrated and measured
value of the sensor is
smaller
than the reference
measurement
Possible cause
Running-in time too short
Diaphragm cap damaged
Short circuit in the
signal cable
Interfering substances
Running-in time too short
Coating/deposits on
the diaphragm cap
No sample water fl ow
Air bubbles outside
on the diaphragm
Harmful substances
in sample water
Coating/deposits (mangan-
ese, iron oxide) at the
diaphragm cap
Remedy
See “Running-in time”
Change diaphragm cap;
run in the sensor, calibrate
Locate the short circuit and
remove
Contact ProMinent
See “Running-in time”
Clean or change diaphragm
cap; run in the sensor,
calibrate
Correct the fl ow
Increase the fl ow within the
permitted levels
Consult ProMinent
Clean or change diaphragm
cap; run in the sensor,
calibrate
BA_DR_002_02_09_GB_Cooling.indd 53
BA_DR_002_02_09_GB_Cooling.indd 53
24.02.2009 11:56:06 Uhr
24.02.2009 11:56:06 Uhr