To determine the precise cell constant (cc) of the sensor, place the sensor in a calibration
so lu tion (sample 1) with a known conductivity and select the second menu option. The
DULCOMETER
®
D1C displays the conductivity it has calculated with the updated parameters.
Press Enter when the value is constant. Select the next menu option and enter the conductivity
of the calibration solution (arrow keys). Press Enter. The D1C displays the recalculated cell
constant and saves the value. The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops. The limit
value monitor and the fault diagnosis system are reset. The standard signal of the “measured
value“ output is frozen.
Possible
values
Initial value
Increment
Lower value
Upper value
Remarks
Conductivity Measured
0.1 μS/cm
0 μS/cm
200 μS/cm
Measuring range
Calibration
value
200
μS/cm
solution (LF)
1 μS/cm
0 μS/cm
2000 μS/cm
Measuring range
2000
μS/cm
0.01 mS/cm
0 μS/cm
20 mS/cm
Measuring range
20
mS/cm
0.1 mS/cm
0 μS/cm
200 mS/cm
Measuring range
200
mS/cm
1 mS/cm
0 μS/cm
2000 mS/cm
Measuring range
2000
μS/cm
Inductive Conductivity
Parameter
Record
This setting menu can be used to temporarily enable additional parameters of the second
parameter record PAR2 to be edited (in addition to the parameters from the setting menu
“measurement parameter setting?“). I.e. the “limit setting?“ and “mA output 1 (2) setting?“ setting
menus are enabled to allow editing of PAR2 (designation PS2) until you exit the series of setting
menus from “parameter record“ to “general setting“.
It is possible to switch between the parameter records for operation using the “feed forward
control“ contact input, e.g. during a fl ushing process (contact open = PAR1, contact closed =
PAR2). When switching between the parameter records, the current parameter record in the
setting menu “measurement parameter setting?”, “limit setting?“ and “mA output 1 (2) setting“ is
replaced by the other parameter record (designation PS1 or PS2).
There is no access to the calibration menus in the case of PAR2 (designation PS2).
NOTE
• The designation PAR1 indicates that the parameter record 1 is actively measuring
(contact input “feed forward control“ e.g. open).
The designation PS1 indicates that the parameter record 1 can be edited.
Exception: The parameter record 1 cannot be edited while it is actively measuring
(designation PAR1). If your controller switches the D1C to parameter set 1 while it is
being edited, the D1C returns to the permanent display.
• PAR2, no control, no standard signal outputs.
Grenzwert2 oben
5.00 ppm
Grenzwert1 unten
1.00 ppm
Only with parameter record 2
D1C2-cond.-042-GB
parameter record
setting ?
PAR1
parameter record
PS1
Possible
values
Initial value
Increment
Lower value
Upper value
Remarks
Parameter
PS1
PS1
record
PS2
BA_DR_002_02_09_GB_Cooling.indd 88
BA_DR_002_02_09_GB_Cooling.indd 88
24.02.2009 11:56:09 Uhr
24.02.2009 11:56:09 Uhr