29
29
0.0 - 4.5 mg/l
(Br )
2
Name on device: 128-Bromine-pp
(128)
Bromine
Powder
Continued...
Note(s)
-> Alkalinity of more than 250 mg/l CaCO affect the measurement and the
3
color development. Neutralize to pH 6-7 with 1 N Hydrochloric Acid before
performing the test.
-> Acidity of more than 150 mg/l CaCO affect the measurement and the color
3
development. Neutralize to pH 6-7 with 1 N Sodium Hydroxide before
performing the test.
-> Other oxidants, such as chlorine, chlorine dioxide, ozone, peroxides as well
as iodine will interfere with the reagent reaction and will be part of the
displayed value.
-> If oxidized manganese or oxidized chromium is present in the sample,
sample needs to be pre-treated.
-> Hardness levels above 1000 mg/l CaCO affect the measurement
3
-> Extreme pH-values of the sample need to be corrected to pH 6-7 before
measuring the sample.
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...