187
187
10 ml
24 mm
4
4
1.) Set up the fridge to 15°C by using the thermostate according to the
manual.
2.) Fill a cleaned cell (24mm) with 10ml of your sample.
3.) Close the lid.
4.) Place the vial inside the fridge for 20:00 minutes.
5.) Place the sealed "Methanol Zero" vial into the PrimeLab and press
ZERO.
6.) Remove the the cell from the PrimeLab and set it aside. It is not longer
required for this test.
7.) Remove the cell from the fridge.
8.) Add 1 PTT Phot. tablet to your sample.
9.) Crush the tablet with a clean stirring rod.
10.) Close the lid.
11.) Place the vial inside the fridge for 10:00 minutes.
12.) Press TEST to start 10:00 minutes countdown.
13.) Remove the cell from the fridge.
14.) Remove condensate on vial by using a dry soft cloth.
15.) Place the cell into the photometer (use caution to the arrow on the
cell) and press TEST.
Tablet
Name on device: 159-PTT-tab
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
(159)
Permanganate Time Test
0 - 100 %A
(Absorption)
PTT Photometer
(TbsPTT...)
Continued...
4
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...