26) Add 1 x 0.05ml (scoop) "PL Phosphonate 4" powder to the sample
water .
27) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
28) After the lapse of a 10 minute countdown the total result is displayed,
divided in:
"tPO " = "Organophosp Phosphate as PO ",
4
4
"PO " = "Phosphate as PO ",
4
4
„Po Org" = "Organophosphonate as PO ".
4
4
Scroll through the results using the buttons "arrow up" and "arrow down".
Press button 3 to convert this into various measurement units (**).
Press button 4 to save the result.
0 - 20 mg/l
Liquid
+Powder
Continued...
4
Name on device: 87-Phosphonate-liq
4
3
(87)
Phosphonate
Note(s)
->
(*) Special accessories required / not included as standard equipment!
->
(**) Conversion to PBTC, NTP, HEDPA, EDTMPA, HMDTMPA,
DETPMPA, HPA
-> With this procedure ortho-phosphate ions are detected. Other phosphates
must therefore be converted to ortho-phosphates before the test is begun.
-> The pH value of the sample water should be between 6 and 7 pH.
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...