Notes
-> Filter sample if neccessary to test a clear sample.
-> Make sure the temperature of your sample does not exceed 30°C.
-> High chlorine levels (> 30 mg/l) interfere with this test and can influence the
reading.
-> The test needs to be performed without a delay. Place the vial into the
PrimeLab right after reagents have dissolved and lid is closed. Immediately
press TEST. It is essential for the accuracy of this test to keep the countdown
of 02:00 minutes right after dissolving the tablets/closing the lid/placing the
vial into the PrimeLab.
->
DO NEVER SHAKE THE VIAL!
-> For expected values below 400 mg/l it is strongly recommended - to achieve
the most accurate result - to better use ID 36 (Nitrite with powder reagents
0 - 200 mg/l; extended range 0 - 400 mg/l by 1:1 dilution).
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Continued...
(97)
Nitrite (HR)
0 - 1500 mg/l
(NaNO )
2
Tablet
Name on device: 97-Nitrite-HR-tab
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...