12.) Fill a clean luer-lock syringe (20 ml) with deionised water.
13.) Connect the syringe to the filter holder and discharge the syringe to
waste.
14.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
15.) Fill a clean luer-lock syringe (20 ml) with deionised water.
16.) Connect the syringe to the filter holder and discharge the syringe to
waste, down to the 10 ml mark.
17.) Filter the remaining 10 ml of destilled water into a clean 24mm vial.
18.) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
19.) Remove the cell and open the lid.
20.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
21.) Connect the syringe to the filter holder and discharge the syringe to
waste.
22.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
23.) Connect the syringe to the filter holder and discharge the syringe to
waste.
24.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
25.) Connect the syringe to the filter holder and discharge the syringe to
waste, down to the 10 ml mark.
26.) Filter the remaining 10 ml of SAMPLE water into a clean 24mm vial.
27.) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press TEST.
4
Name on device: 107-Colour
(107)
Colour
Continued...
Continued...
4
0 - 500 mg/l
(Pt-Co)
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...