Liquid
Note(s)
-> (*) Special accessories required / not included as standard equipment!
-> If the sample needs to be filtered (dissolved iron) it must be ensured that
after filtration there are no more undissolved iron parts in the sample. If you
are unsure, please repeat the filtration.
-> High nitrite values in the sample water can influence the measurement. If the
sample water turns to red or pink after adding "PL Iron HR 1" drops a new
sample needs to be taken with 0.1 g "TN1" powder added to it. After adding
the powder wait for 2 minutes and then start the measurement procedure as
described on the front page.
(30)
Iron (HR)
Name on device: 30-Iron-HR-liq
0 - 30 mg/l
2+/3+
(Fe
)
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
8) Add 15 drops "PL Iron HR 1" liquid reagent to the test liquid in the
cuvette.
9) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
10) After the lapse of a 5 minute countdown the determined result is
displayed. Press button 4 to save this together with date and time in the
device.
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...