67
67
0 - 500 mg/l
(Pt-Co)
Name on device: 107-Colour
10 ml
24 mm
(107)
Colour
Select either:
A) „True colour“
B) „Apparent colour“
A) Measurement procedure for „True colour“
1.) Unscrew the two halves of the filter holder and place one filter circle
(0.45µ) onto the base section.
2.) Screw the two parts together again, ensuring the O ring is correctly
located.
3.) Fill a clean luer-lock syringe (20 ml) with deionised water.
4.) Connect the syringe to the filter holder and discharge the syringe to
waste.
5.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
6.) Fill a clean luer-lock syringe (20 ml) with deionised water.
7.) Connect the syringe to the filter holder and discharge the syringe to
waste.
8.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
9.) Fill a clean luer-lock syringe (20 ml) with deionised water.
10.) Connect the syringe to the filter holder and discharge the syringe to
waste.
11.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
(*)
Syringe
10
9
8
7
6
5
4
3
2
1
ml
Filter
0.45µm
(*)
Filter paper
(*)
Filter adapter
Continued...
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...