217
217
1) Add 5ml sample water into a new and clean Phosphorus LR cell.
2) Add 2 x 0.5ml scoops of PL Phosphorus 2.
3) Immediately close the lid and shake vial vigorously for 00:20 minutes.
4) Press TEST to start 00:20 minutes countdown.
5) Put the cells for 30:00 minutes at 150°C in a preheated thermoreactor.
6) Press TEST to start 30:00 minutes countdown.
7) CAUTION: cells are hot! Remove the cells from the thermoreactor and
allow to cool to 60°C or less.
8) Add 10 drops PL Phosphorus LR1 to your sample. Shake the cell to mix
the liquid.
9) Place the COD adaptor into the device.
10) Do you want to use the last stored ZERO value?
11) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
12) Remove the cell and open the lid.
(153)
Phosphorus-Total (LR)
0.00 - 2.60 mg/l
(P)
Name on device: 153-PsphrTotLR-tab
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
16mm vial
Phosphorus LR vial
PL Phosphorus LR1
PL Phosphorus 2
Phosphate LR 1
Phosphate LR 2
Continued...
4
4
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 10)
-> Use last ZERO
= continue with step 13)
1
2
3
4
4
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...