225
225
(111)
PTSA
0 - 1000 µg/l
(PTSA)
Do ONLY use the cuvette which has been used to do calibration
for this parameter!
Use 10ml Pipette to always properly dose exactly 10ml!
1) Fill sample water into the same cuvette up to the 10ml level.
2) Insert the cuvette WITHOUT LID into the PrimeLab sampling chamber.
Align the arrow on the front of the cuvette.
3) Place PTSA-Adapter on top of the open cuvette which already is placed
into the PrimeLab measurement chamber.
4) Press TEST.
5) The determined result is immediately displayed. Press button 3 to
convert the result to different units (*). Press button 4 to save the result
together with date and time.
10 ml
Name on device: 111-PTSA-Ad
24 mm
4
Adapter
PLSp-PIP10ml
PLSp-PIP10ml-tips
PLSp-ADP-PTSA
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
3
4
PLSp-RefPTSA
Only for calibration!
Continued...
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...