103
103
0 - 500 mg/l
(CaCO )
3
1) Fill a cleaned cell (24mm) with 10 ml of your sample.
2) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
3) Remove the cell and open the lid.
4) Add 10 drops POL20TH1 to your sample. Shake the cell to mix the
liquid.
5) Close the lid. Shake the cell gently to mix the liquid.
6) Add 4 drops POL10TH2 to your sample. Shake the cell to mix the liquid.
7) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press TEST.
8) Time left:
02:00 minutes.
Continued...
(148)
Hardness-total (HR)
Name on device: 148-Hard-tot-liq
10 ml
24 mm
10
9
8
7
6
5
4
3
2
1
ml
4
4
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
10 ml
24 mm
10
9
8
7
6
5
4
3
2
1
ml
Liquid
PL Total Hardness 1
PL Total Hardness
2
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...