273
273
(62)
Zinc
Tablet
0 - 1 mg/l
(Zn)
Copper/Zinc LR Photometer
(TbsRCZ...)
EDTA
(TbsHED...)
10 ml
Name on device: 62-Zinc-tab
24 mm
1) Please select Chlorine Dioxide:
A:
"in presence of chlorine“
B:
"in absence of chlorine“
A)
"in presence of chlorine“:
2) Fill a cleaned cell (24mm) with 10ml of your sample.
3) Do you want to use the last stored ZERO value?
4) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
5) Remove the cell and open the lid.
6) Add 1 Dechlor tablet to your sample.
7) Crush the tablet with a clean stirring rod until the tablet has completely
dissolved.
8) Add 1 Copper/Zink LR tablet to your sample.
9) Crush the tablet with a clean stirring rod until the tablet has completely
dissolved.
10) Close the lid. Shake the cell gently to mix the liquid.
11) Place the cell into the device. Again, please use caution to the arrow
on the cell. Than press TEST.
12) Remove the cell and open the lid.
13) Add 1 EDTA tablet to your sample.
14) Crush the tablet with a clean stirring rod until the tablet has completely
dissolved.
15) Close the lid. Shake the cell gently to mix the liquid.
16) Place the cell into the device. Again, please use caution to the arrow
on the cell. Than press TEST.
Continued...
4
4
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 4)
-> Use last ZERO
= continue with step 6)
1
2
3
4
4
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...