Menu: SET (Setup)
SET 5
SET 5
Continued...
Calibration
The following refers to calibration of ID 111/112/113 (PTSA/Turbidity/Fluorescein):
As this Parameter uses an indirect light from above, it is essential to always ensure the
correct amount of water in the vial, which is why exactly 10 ml of liquid should be taken
when doing a test. To achieve this, use the pipette which is part of the Adapter-kit. Please
change or clean the tip of the pipette after each measurement/calibration using destilled
water.
Please perform a new calibration for the parameters of the ID111 / 112 / 113 if the
measurement results obtained appear to be inaccurate or no longer correspond to
the value of the standard solutions supplied.
If you select „Turbidity-adapter“:
You first have to perform a „PrimeLab calibration“ !!! See page „SET 4"
Calibration steps, as shown below, will be displayd on the PrimeLab as well. Please
make sure the standard vials (0.5/10/1000 NTU) are free from fingerprints and dirt.
Clean vials properly by using the cloth provided in your kit, prior inserting into the
PrimeLab.
Use caution to shaking-/rest-instructions
on the standard vials (0.5/10/1000 NTU):
0.5 NTU
standard vial: Let vial rest for at least 3 hours before use for the calibration
purposes. Turn vial
TWO TIMES
before you open it.
Do not shake!
10 NTU
standard vial: Turn vial
THREE TIMES
before you open it.
Do not shake!
1000 NTU
standard vial:
Shake vial heavily
before you open it!
Open „
0.5 NTU
“ standard vial and insert
without lid
and with forward arrow into the
PrimeLab.
Set Turbidity adapter on top of the
open cell.
Press TEST.
Remove vial from PrimeLab and close it properly for future use.
Open „
10 NTU
“ standard vial and insert
without lid
and with forward arrow into the
PrimeLab.
Set Turbidity adapter on top of the
open cell.
Press TEST.
Remove vial from PrimeLab and close it properly for future use.
Open „
1000 NTU
“ standard vial and insert
without lid
and with forward arrow into
the PrimeLab.
Set Turbidity adapter on top of the
open cell.
Press TEST.
ID 112 is now calibrated successfully.
4
4
4
Continued...
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...