Pressure Systems, Inc. Model 9116 User’s Manual
Page 64
www.PressureSystems.com
Example:
●
Send TCP/IP command to a
Model 9116
module (via its open socket) to
calculate and set new offset coefficients for channels 16 through 13.
“
hF000
”
Read response, containing all new offset values (also stored in the module’s
volatile main memory):
“
0.0010 0.0020 0.0015 0.0025
”
Actual offset values are returned in the above response as decimal fixed-point
ASCII strings, each preceded by a
space
character. From left-to-right: they are
for channels 16, 15, 14, and 13.