Pressure Systems, Inc. Model 9116 User’s Manual
Page 30
www.PressureSystems.com
RESET (Command ‘B’)
Purpose:
Instructs the module to reset internal operating parameters, and to set all internal
control variables to their default “reset” state (see description below). The current
TCP/IP socket connection will remain open. Execution after a power off/on cycle
is optional (unnecessary).
Command
“
B
”
‘B’ is the
command
letter.
Response
“A”
‘A’ is the
acknowledge
letter.
Description:
The module returns to the following “reset” states if this command is executed:
●
Re-zero correction (offset) terms are set to the last values stored in
transducer
memory.
●
Span correction (gain) terms are set to the last values stored in transducer
memory.
●
Calibration Valve is set to the RUN Position
●
Number of Samples for Data Averaging is set to last value stored in non-
volatile memory (factory default = 8).
●
Any
autonomous
host data delivery
streams
defined by
‘c’
sub-commands
are
reset
(undefined).
●
The
Multi-Point Calibration
function defined by
‘C’
sub-commands is reset
(undefined) if in progress.
Example:
●
Send TCP/IP command to a
Model 9116
module (via open socket) to reset defaults:
“
B
”
Read following response:
“
A
”