Example:
●
Send TCP/IP command to
Model 9116
module (via its connected socket) that returns
decimal raw “pressure” A/D counts data for channels 1, 5, 9, and 13:
“
a11110
”
Response contains data for channels 13, 9, 5, and 1 (left to right):
“
32767.000000 -32700.000000 10.000000 16385.000000
”
Please note that channel 13 is saturated at +full scale and channel 9 is almost saturated
at -full scale. Channel 5 reads near zero and channel 1 is about ½ +full-scale.