Pressure Systems, Inc. Model 9116 User’s Manual
Page 70
www.PressureSystems.com
06
IP Address Resolution Method
, as hex state: (default = 0000)
0000
=
Use Static IP Address
stored in module’s non-volatile
memory
0001
=
Get Dynamic IP Address
from external RARP/BOOTP
server
13
07
Host Response/Stream Back-Off Delay
, as hex
value
(or FFFF). FFFF
means use low-order byte of module’s Ethernet Address as
value
instead. In either case,
back-off delay
in
microseconds
is calculated from
decimal
equivalent of hex
value
:
delay =
decvalue
* 20
14
08
Host Response/Stream Total Size Prefix
(with 2-byte big-endian binary
value), added to all command responses and streams to indicate their
true length in bytes:
0000 = None (default)
0001 = Yes
16
09
TCP Connect Port
, as hex value (e.g. 2328 = 9000
decimal
, default).
17
0A
Auto UDP Broadcast@Reset
, as hex state:
0000 = No (default)
0001 = Yes
18
0C
Temperature Status of Each Scanner Transducer
, as 16-bit hex bit
map, each bit representing the current status of a transducer/channel
(16-1)). Bit values are:
0= transducer operating within the specified operational limits.
1= transducer operating outside the specified limits.
(see end-of-table NOTE +)
0D
Minimum Temperature Alarm Set Point
(in degrees C), as decimal (**)
format 0 representation of internal IEEE float, with leading
space
).
19
0E
Maximum Temperature Alarm Set Point
(in degrees C), as decimal (**)
format 0 representation of internal IEEE float, with leading
space
).
19
11
Thermal Update Scan Interval
(in seconds) as decimal (**) integer
value.
1B
31
Module hardware version
number in the form e.eeeeee. (9016 returns
‘N08’
32
Hardware trigger mode.
0=positive
going
edge
1=negative
going
edge
2=trigger
on
any
edge
(9016
returns
a
‘N08’)
32
3c
Temperature Range
, as a hex value
0000 = range 0 to 60ºC (default)
0006 = range -30 to 60º C
0007 = range -20 to 70ºC
(+)
NOTE
:
This 4-byte hex status fields may also be returned in autonomous data streams, but as
pure binary extensions of each stream’s packet binary header (see
‘c’
command,
ii
=05,
bbbb
=0002).