Pressure Systems, Inc. Model 9116 User’s Manual
Page 38
www.PressureSystems.com
Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration
This sub-command has no additional parameters.
Command
“
C 03
”
‘C’ is the
command
letter.
‘ 03’ is the
sub-command index
(
ii
) for
Abort
.
NOTE: all parameters are separated by a
space
.
Response
“
A
”
‘A’ is the
acknowledge
letter
Description:
Aborts the
Multi-Point Calibration
function, if it is currently in progress. This
sub-command also restores the original “A/D samples for averaging” count to the
module that was in use before the calibration function was started.
It should be noted that executing the
Configure & Start
(‘00’) sub-command
again, after the calibration function has started collecting data (per
Collect Data
(‘01’) sub-commands), but before the final data are calculated (per
Calculate &
Apply
(‘02’) sub-command), will have the same affect as this Abort function.
Example:
●
Abort the Multi-Point Calibration function previously started
“
C 03
”
Read
response:
“A”