Pressure Systems, Inc. Model 9116 User’s Manual
Page 42
www.PressureSystems.com
Internal firmware limits calculated gains to values are software limited to values between 0.0
and 100.0. Any calculated value outside of this range will result in the gain coefficient being set
to 1.00.
Note
The calculated
gain
values from the latest ‘Z’ command will be lost
when the module is powered off. To save these
gain
terms to each
transducer’s non-volatile memory, refer to the
Set Operating Options
(‘w’) command (index 09).
Example:
●
Send TCP/IP command to a
Model 9116
module (via its open socket) to
calculate and set gain coefficients for channels 8 through 4. Instruct the module
to use 14.8890 psi as the applied pressure instead of each transducer’s full-scale
value:
“
Z00F8 14.8890
”
Read response, containing the new gain values (also stored in the module’s
volatile main memory):
“
1.000212 1.000269 1.000437 1.000145 .999670
”
Actual gain values are returned in the above response as decimal ASCII strings,
each preceded by a
space
character. From left-to-right: they are for channels 8,
7, 6, 5, and 4.