Note
Coefficients used for typical applications are shown in BOLD type.
All other coefficients are typically not used outside of advanced
diagnostic functions.
cc
Transducer Coefficients Description
Datum
Type
00
Re-zero Cal Adjustment
(offset) term
FLOAT
01
Span Cal Adjustment
(gain) term
FLOAT
02
Dynamic EU Conversion coefficient c0
FLOAT
03
Dynamic EU Conversion coefficient c1
FLOAT
04
Dynamic EU Conversion coefficient c2
FLOAT
05
Dynamic EU Conversion coefficient c3
FLOAT
06
Reserved for Factory Use
---
07
User Defined Date field (see end-of-table note)
INTEGER
08
Date of Factory Calibration (see end-of-table note)
INTEGER
09
Transducer Manufacturing Reference number
INTEGER
0A
Transducer Full-Scale Range code (see Appendix F)
INTEGER
0B-0F Temperature 1, Pressures 1-5 voltages
FLOAT
10-14 Temperature 2, Pressures 1-5 voltages
FLOAT
15-19 Temperature 3, Pressures 1-5 voltages
FLOAT
1A-1E Temperature 3, Pressures 1-5 voltages
FLOAT
1F-23 Temperature 5, Pressures 1-5 voltages
FLOAT
24-28 Temperature 6, Pressures 1-5 voltages
FLOAT
29-2D reserved for future use (temperature 7)
FLOAT
2E
Temperature 1 Temperature Output voltage at 0 psi
FLOAT
2F
Temperature 2 Temperature Output voltage at 0 psi
FLOAT
30
Temperature 3 Temperature Output voltage at 0 psi
FLOAT
31
Temperature 4 Temperature Output voltage at 0 psi
FLOAT
32
Temperature 5 Temperature Output voltage at 0 psi
FLOAT