Pressure Systems, Inc. Model 9116 User’s Manual
Page 85
www.PressureSystems.com
ii
dd
Description
‘q’ read
index
31
00
Set module type alias.
1
eeee=9116 (default) or 9016
00
32
dd
Set hardware trigger mode.
2
dd=00 trigger on positive going edge (default)
dd=01 trigger on negative going edge
dd=02 trigger on any edge (duty cycle must be taken into
account in order to avoid over-triggering)
32
3c
00
06
07
Set temperature range = 0 to 60ºC (default)
Set temperature range = -30 to 60ºC
Set temperature range = -20 to 70ºC
1.
The set module type alias command instructs the 9116 to identify itself as a 9016 in response to the ‘psi9000' UDP broadcast
command, the UDP startup broadcast messages, and in response to the q0 command. This command is provided in support
of legacy software. Valid command responses are; ‘A’ - acknowledge, ‘N08’ - invalid command (9016s always return’N08’),
and ‘N07’ - invalid alias type. The module type alias setting can be made nonvolatile with the ‘w07’ command.
2. The Set hardware trigger mode is always volatile and will be set to the default upon any interruption of power. Valid command
responses are; ‘A’ - acknowledge and ‘N08’ - invalid command (9016s always return ‘N08’).
Modification of option 13 hex results in the new option selection becoming the module’s new
power-on default. All other options must be stored in non-volatile flash memory using ‘
w07
’
command in order to be retained after the module power cycles.