Pressure Systems, Inc. Model 9116 User’s Manual
Page 35
www.PressureSystems.com
Example:
●
Supply each of the previously-specified three (3) pressure calibration points to the Multi-
Point Calibration function, as was stated in the
previous example
of the
Configure and
Start
(‘00’) sub-command. Assume that all the affected four (4) channels have
differential transducers with the same -5 to +5 psi range. Include at least one pressure
point in the negative range of these transducers
“
C 01 1 -2.5
”
“
C 01 2 0.0
”
“
C 01 3 5.0
”
Read responses (separately after each command executed above):
“-2.4998 -2.4999 -2.5001 -2.500”
“0.0 0.0013 -0.0133 -0.00001”
“5.0091 4.9992 5.0010 4.9998”
Data are returned in
reverse
channel number order (i.e., 4, 3, 2, 1) in each response.