60 .
LS 55 User’s Guide
10.
Start a TimeDrive using suitable wavelengths for the fluorescent dye used,
using slit widths of 10 nm (excitation) and 5 nm (emission).
Duration should be 2-3 minutes.
11.
During data collection, adjust the two (1/8") Hex screws shown in
Figure 41 until the maximum signal is observed.
Figure 41 Adjustment screws
12.
Remove the accessory from the LS 55 sample area and tighten the locking
screws shown in Figure 39.
13.
Re-fit the front plate, as shown in Figure 38.
14.
Re-insert and re-clamp the tubing using the tubing clamp screw, as shown in
Figure 38.
Operation
The Sipper accessory can be controlled manually using the LS 55 Status application.
The Scan and Concentration applications include full sipper control. The Sipper
parameters (pump time, wait time, purge time and direction) are set in the
application so that when the Sipper is started by pushing the Sipper arm upwards,
sampling and measurement are automatically triggered.
Manual control
1.
Start the
LS 55 Status
application from the FL WinLab software.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...