Accessories: Installation and Use . 43
Screw-fitting Flowcell
Figure 21 Screw-fitting flowcell
Description
The screw-fitting flowcell (B0631133) is intended for automated flow-through, for
example using an autosampler.
NOTE:
The flowcell is not intended for HPLC, or other high resolution applications
(see instead the
on page 63).
The flowcell has three main advantages:
•
The screw fittings allow tight connections to be made with autosamplers or
automated flow systems where high sample throughput requires high reliability
due to the unattended nature of data collection. Use the PerkinElmer Flange
toolkit, Part number B3000001.
•
The internal volume of the flowcell (the internal chamber has dimensions
3x3x10 mm, a volume of 90
μ
L) is almost perfectly matched to the light beam
geometry of the LS 55 (the measured volume is 3x3x9 mm, a volume of 81
μ
L).
This means that this flowcell does not suffer from the sensitivity losses
normally associated with flowcells, but gives excellent sensitivity, equivalent to
a normal cuvette.
•
The flowcell fits into a standard cuvette holder, allowing the flowcell to be
thermostatted (as detailed in
Installing Thermostatting Tubing to Cellholders
on
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...