RS232 Commands . 139
TLC Scan Commands
Command:
TLC DATA INTERVAL
Syntax:
$IT interval
Description:
Sets data output interval for subsequent TLC scans. Interval is in
range 0.1 to 2 mm and is rounded to nearest multiple of
0.1 mm. Power-on default is 0.5 mm.
Command:
TLC X AXIS-HGH LIMIT
Syntax:
$XH limit
Description:
Sets high limit for TLC x scan. Range 0.1 to 99.0 mm. High limit
may not be lower than TLC x low limit. Power-on default is 63.0
mm.
Command:
TLC X AXIS LOW LIMIT
Syntax:
$XL limit
Description:
Sets low limit for TLC x scan. Range 0 to 98.9 mm. Low limit
may not be higher than TLC x high limit. Power-on default is 0
mm.
Command:
TLC Y AXIS HIGH
Syntax:
$YH limit
Description:
Set high limit for TLC y scan. Range 0.1 to 63.0 mm. High limit
may not be lower than TLC y low limit. Power-on default is 63.0
mm.
Command:
TLC Y AMS LOW
Syntax:
$YL limit
Description:
Set low limit for TLC y scan. Range 0 to 62.9 mm. Low limit may
not be higher than TLC y high limit. Power-on default is 0 mm.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...