RS232 Commands . 161
Fast Filter Accessory (FFA) Commands
Command:
SET EXCITATION UNIT STATE
Syntax:
$FX state
Description:
Sets excitation unit state if fitted. Valid states are: 0 clear beam
(power-on default), 1 position 1, 2 position 2, 3 position 3, 4
position 4, 8 running synchronous. If no parameter is specified,
the current excitation unit state is returned.
Command:
SET EMISSION UNIT STATE
Syntax:
$FM state
Description:
Sets emission unit state if fitted. Valid states are as for $FX
command. If no parameter is specified, the current excitation
unit state is returned. Error code 0100 means units are not fitted.
Error code 0112 signifies a synchronization error.
NOTE:
Excitation and emission units may not be run simultaneously.
Command:
SET FFA CYCLE LENGTH
Syntax:
$FI time
Description:
Sets FFA data output cycle length rounded up to next multiple of
4 mains cycles from input time; time range 0.00 - 9.99s.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...