Flowcells are marked with an arrow that indicates the inlet side, as the reagent
stream on the inlet side is brought to the bottom of the flowcell, assisting in de-
bubbling.
This inlet side should be routed to the sample tube, or to the sampling needle of
the autosampler. The outlet side should be routed to the peristaltic pump. An
example is shown in Figure 22.
Figure 22 Connecting the flowcell
The system should be cleaned regularly with the solvent used in the analysis. If
the system is to be left overnight or longer without being used, then the system
should ideally be cleaned with de-ionized water, then with ethanol, then pumped to
dryness.
Operation
The inlet and outlet tubing should be installed with screw fittings using the
PerkinElmer Flange tool (B3000001). This enables you to fit a wide range of screw
fittings to various diameter tubing. Ideally, PTFE or FEP tubing should be used, as
this is chemically inert.
NOTE:
The internal diameter of the tubing is important, too small a diameter could
lead to inefficient pumping and cavitation, producing air bubbles. Too large
a diameter will lead to large sample volumes. Typical internal diameter is
0.7-1.0 mm.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...