Accessories: Installation and Use . 45
Installing Thermostatting Tubing to Cellholders
All cuvette holders supplied for use with the LS 55 are thermostattable using
circulating water:
•
The cuvette holders are made watertight with a sealant that is guaranteed up
to 60
°
C.
•
Higher temperatures may make the sealant leak, in which case you should
replace it with a suitable sealant with higher temperature specification.
•
At temperatures below 15
°
C, condensation may occur on the surfaces of the
cuvette. If this is the case, either purge the sample area with dry air, or lead
dry air onto the cuvette surfaces using tubing.
To fit the thermostatting tubing:
1.
Route the tubing into the sample area and onto the two tubing ports under
the sample area door, as shown in Figure 23.
Figure 23 Tubing ports
2.
The tubing is then brought around the outside of the accessory bracket, as
shown in Figure 24.
Tubing ports
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...