Accessories: Installation and Use . 49
2.
Check that the cap fits securely.
If not, the septum is probably too thick to allow the cap to grip the barrel
adequately, and should be replaced.
3.
Remove the blanking plug from the LS 55 sample area door, as shown in
Figure 29.
Figure 29 The blanking plug
4.
Insert the septum injector.
NOTE:
The distance between the top of the cuvette and the bottom of the injector
barrel avoids the thermostatting tubing fouling the injector when the door is
closed. For this reason, a short syringe needle will not reach the cuvette,
and there is no guarantee that injected reagents will enter the cuvette. A
minimum syringe needle length of 70 mm is required to ensure that the
needle enters the cuvette.
Blanking plug