![PerkinElmer LS 55 User Manual Download Page 184](http://html1.mh-extra.com/html/perkinelmer/ls-55/ls-55_user-manual_1530217184.webp)
184 .
LS 55 User’s Guide
CAUTION
Excessive force will break the lamp, leaving sharp pieces in the
lamp housing.
9.
Remove the new lamp from its packaging.
CAUTION
Take care-the lamp is EXTREMELY sensitive. A slight knock
against the table surface, for example, can immediately break
the lamp. Also take care not to touch the lamp, or the central
capillary section could become contaminated leading to serious
loss of sensitivity. If the capillary is accidentally touched, wipe
immediately with spectroscopic grade ethanol or methanol and
dry with a cuvette-wipe tissue.
NOTE:
In order to seat the lamp correctly in the bracket, it needs to be rotated
before insertion into the bracket so that it is correctly aligned with the
optical system.
10.
Turn the lamp so that the trigger strip, as shown in Figure 127, is located to
the right, pointing towards the small mirror next to the entrance slit, as
shown in Figure 128.
Figure 127 Details of the electrode
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...