6.
Attach the excitation filter wheel control cable, as shown in Figure 102.
Figure 102 Attaching the excitation control cable
7.
Fix the emission polarizer into the emission filter wheel location, using the
locating hole and pin for initial location of the wheel, as shown in
Figure 103.
Figure 103 Fixing the emission polarizer wheel in position
8.
Before tightening the fixing screws fully, check that the emission filter wheel
and emission polarizer wheel are correctly set and do not touch one another
during rotation, as shown in Figure 104.
Do this by manually rotating the two wheels in opposite directions. If the
wheels touch, gently turn the polarizer wheel until the two wheels do not
meet.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...