LS 55 User’s Guide
Command:
INITIALIZE CELL HOLDER
Syntax:
$CI
Description:
Drives cell holder to datum position. Used after any manual
change to cell holder accessory.
Command:
MOVE CELL HOLDER
Syntax:
$CM position
Description:
Moves cell holder to the specified position. Position is an integer
in the range 1 to 4. Initialization default is 1.
Command:
PERISTALTIC PUMP BLOW
Syntax:
$PB time
Description:
Causes the peristaltic pump to blow for the specified time. Time
is in the range 0.1 to 100 s and is rounded to 0.1 s
Command:
PERISTALTIC PUMP SUCK
Syntax:
$PS time
Description:
Causes peristaltic pump to suck for the specified time. Time is in
the range 0.1 to 100 seconds and is rounded to nearest multiple
of 0.1 s. Power-on default is 0.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...