RS232 Commands . 133
Monochromator Scan Control Commands
Command:
ABSCISSA HIGH LIMIT
Syntax:
$AH <abscissa limit>
Description:
Sets either a new emission or excitation high abscissa limit
depending on monochromator selection. Emission limit must be in
range 200.0 to 900.0. Excitation limit must be in range 200.0 to
800.0. Limits are rounded to nearest 0.1 and are in units of wn.
A high limit must be greater than the corresponding low limit.
Power-on defaults for high limits are: 800 for excitation and 900
for emission. Returns high limit for selected monochromator if no
parameter is specified.
NOTE:
Excitation limit is set if monochromator selection is 'both'. Refer to
command $MX.
Command:
ABSCISSA LOW LIMIT
Syntax:
$AL <abscissa limit>
Description:
As for high limit. Power-on defaults for low limits are 200 for
both. Returns low limit for selected monochromator if parameter
is not specified.
Command:
MONOCHROMATOR SELECTION QUALIFICATION
Syntax:
$MQ qualifier
Description:
Causes subsequent scans of both monochromators to be
'constant wavelength' if qualifier = 0, or 'constant frequency' if
qualifier =1. Qualifier must be integer. Power-on default is 0.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...