LS 55 User’s Guide
Command:
MOVE TO WELL NUMBER
Syntax:
$WP well no.
Description:
Moves plate reader to position corresponding to well no. Well no.
is an integer in the range 1 to 96.
Fast Kinetics Application (FKA) Commands
Command:
FKA WAVELENGTHS 1
Syntax:
$Wl ex wavelength, em wavelength
Description:
Sets wavelengths for first output data point of FKA run (and
subsequent alternate data points). If no parameters are given,
outputs current setting (power-on default 340 and 510 nm). If
only one parameter given, assumes other wavelength to be
current monochromator position.
Command:
FKA CYCLE TIME
Syntax:
$FC time
Description:
Sets FKA cycle time (interval between successive points output)
for FKA runs. Time format xxx.x (seconds). Zero time sets
minimum possible cycle time for current $RD integration time,
data collection cycle length, and wavelengths $W1 and $W2
Power-on default 2.1 s. Minimum value for $RD 0.1 and default
W1, W2 is 1.6 s. Returns error 0009 if requested time is smaller
than current minimum possible time. Returns current cycle time if
no parameter is given.
Summary of Contents for LS 55
Page 1: ...LS 55 User s Guide ...
Page 7: ...Warnings and Safety Information ...
Page 12: ...12 LS 55 User s Guide ...
Page 13: ...Introduction ...
Page 19: ...Introduction 19 Supply Voltages 100 V 120 V 220 V 230 V Frequency 50 60 Hz Power 150 VA ...
Page 20: ...20 LS 55 User s Guide ...
Page 21: ...Unpacking and Installation ...
Page 36: ...36 LS 55 User s Guide ...
Page 37: ...Accessories Installation and Use ...
Page 72: ...72 LS 55 User s Guide Figure 54 Using the powder accessory ...
Page 93: ...Accessories Installation and Use 93 Figure 86 Masking the outer areas ...
Page 121: ...RS232 Commands ...
Page 163: ...System Description and Maintenance ...
Page 165: ...System Description and Maintenance 165 Figure 112 The optical layout ...
Page 198: ...198 LS 55 User s Guide ...
Page 199: ...Index ...