IVIS
®
Lumina XRMS Series III Hardware Manual
Chapter 8 | Fluorescence Module
67
8.7 Care and Maintenance of the Fluorescence Equipment
Cleaning the Fluorescence Light Source Module
If necessary, wipe the exterior surfaces of the light source module with a soft cloth.
If the fluorescence light source module requires more aggressive cleaning or sterilization, contact
PerkinElmer Technical Support (see
).
Cleaning the IVIS Lumina XRMS Series III and Fluorescence Equipment
See
Cleaning the IVIS Lumina XRMS Series III
for instructions on cleaning the imaging
chamber,
Cleaning the Optical Components and Filter Replacement
Contact PerkinElmer Technical Support for information about cleaning or sterilizing any of the
optical components or the optical filter replacement (see
!
WARNING!
DO NOT use fluids to clean the exterior or interior of the IVIS Lumina
XRMS Series III. Do not allow fluids of any kind to enter the light source module
under any circumstances. Sprays and liquids that come into contact with the light
source module or the instrument may result in damage to the system or
electrocution.
!
WARNING!
DO NOT use fluids or moistened towels to clean the any part of the
IVIS Lumina XRMS Series III where electrical or fiber optic cables make
connections. Do not use fluids of any kind in the vicinity of the Excitation Filter
Wheel Assembly (mounted on the rear of the imaging chamber). Turn off
electrical power to the instrument before engaging in cleaning operations using
fluids. The Imaging Chamber power switch is located in the rear on the
electronics tray.