IVIS
®
Lumina XRMS Series III Hardware Manual
Chapter 5 | X-ray Safety and Radiation Hazards
24
Additionally, it is recommended to perform a daily or weekly light leak check as described in
. Using the High Reflectance Hemisphere to look for possible light leaks into the
imaging chamber may help detect potential radiation leaks out of the chamber. Performing this check
DOES NOT invalidate the requirement for the 12 month TOTAL radiation survey. It is possible for
X-rays to leak from areas outside of the imaging chamber, for example from the electronics
enclosure.
Regulatory Compliance
Customers in the US are directed to check with their state radiation control program director for
registration requirements. See the Safe Operating and Emergency Procedures for the Operation of the
IVIS Lumina XRMS Cabinet X-Ray System for a list of US state and Canadian Province agencies.
This document was provided as part of the pre-installation instructions and is also included on the
same CD-ROM as this manual. International customers should check with their governing bodies
about possible registration or other requirements. PerkinElmer certifies that the IVIS Lumina XRMS
Series III meets the Japanese standard JAIMAS0101- 2001 with respect to X-ray containment.
PerkinElmer certifies that the IVIS Lumina XRMS Series III complies with FDA regulation CFR
1020.40 after installation at the customer's site. PerkinElmer also certifies that the instrument meets
all of the international regulations of the country where it is installed.
5.6 Contact Information
PerkinElmer Sales and Technical Support
PerkinElmer Health Sciences
940 Winter Street, Waltham, Massachusetts 02451 USA
Telephone: 800.762.4000 (US) or +1.203.925.4602
Fax: +1.203.944.4904
For sales:
For Technical Support:
877.522.497.3302 (US) or +1.508.435.9500
www.PerkinElmer.com
US State Radiation Authorities
ALABAMA
Office of Radiation Control
State Dept. of Public Health
201 Monroe St/PO Box 303017
Montgomery, AL 36130-3017
334-206-5391
http://www.adph.org/radiation/