9
Care and Maintenance
Surveying the IVIS Lumina XRMS Series III for Radiation Leakage
Cleaning the IVIS Lumina XRMS Series III
Cleaning the Scintillation Plate Window and Holder
9.1 Surveying the IVIS Lumina XRMS Series III for Radiation Leakage
PerkinElmer recommends, and some local government agencies may require, an X-ray leakage safety
test be performed:
Every 12 months.
When the system is installed at a new site, unless the instrument was moved using the XWS-260
workbench and no abnormal mechanical shocks occurred.
After PerkinElmer performs maintenance or service.
After any abnormal condition that could impair any of the safety systems. For example, the light
box door becomes difficult to open or close.
Conducting the X-Ray Radiation Survey
A radiation leakage test is complex and requires sensitive equipment. Some states or localities may
require special training and certification to perform the test. Contact PerkinElmer Technical Support
(see
) for information regarding these tests or for scheduling a PerkinElmer representative to
conduct the survey as part of an overall safety check.
9.2 Maintenance and Safety Checks
Daily Safety Checks
The following safety checks should be performed on a daily basis.
1.
Verify that the door interlock in good repair (see
).
2.
Verify that the key switch functions properly.
3.
Verify that the "X-Ray On" red indicator lights are functioning properly when the X-ray modality
is used. The lights are located:
On the lower electronics panel.
In the camera cover.
On the rear of the instrument (only visible from the rear of the instrument and is intended to
warn service personnel).
4.
Verify that the amber "X-ray armed" indicator is working.
5.
Light leak check described in
.