IVIS
®
Lumina XRMS Series III Hardware Manual
Appendix A | XWS-260 Workstation
79
8.
Allow the system to initialize.
You will hear the motors move. The System Status box displays the current changes.
The temperature square in the IVIS Acquisition Control Panel is red at startup and turns green
when the operating temperature is reached. The control panel displays the current temperature
(
The instrument is ready for operation after the temperature is locked at -90°C, as indicated by the
green light in the control panel (
. For operating instructions, see the Living Image
®
Software Manual for IVIS Lumina XRMS Series III.)
9.
To use the X-ray modality:
a.
Verify that the Emergency OFF switch is in the ON position.
b.
Turn the key selector switch to the ON position.
c.
Confirm that the amber arming switch has been pushed and the light is ON.
d.
Verify that the X-ray source is armed.
e.
Use the IVIS Acquisition Control Panel in the Living Imaging software to turn on the X-ray
source during an imaging sequence or session.
Figure A.2
IVIS Acquisition Control panel – Living Image software
The temperature square is green when
the operating temperature is reached.
Click the temperature square to
display the current temperature.